750 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:FLEXray:AUTosetup
(see page 1276)
Command Syntax
:SBUS<n>:FLEXray:AUTosetup
The :SBUS<n>:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
Sets the selected source channel's impedance to 50 Ohms.
Sets the selected source channel's probe attenuation to 10:1.
Sets the trigger level (on the selected source channel) to -300 mV.
Turns on trigger Noise Reject.
Turns on Serial Decode.
Sets the trigger to the specified serial bus (n of SBUS<n>).
See Also "Introduction to :TRIGger Commands" on page 1007
":SBUS<n>:FLEXray:TRIGger" on page 758
":SBUS<n>:FLEXray:BAUDrate" on page 751
":TRIGger[:EDGE]:LEVel" on page 1033
":SBUS<n>:FLEXray:SOURce" on page 757