750 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:FLEXray:AUTosetup
(see page 1276)
Command Syntax
:SBUS<n>:FLEXray:AUTosetup
The :SBUS<n>:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
• Sets the selected source channel's impedance to 50 Ohms.
• Sets the selected source channel's probe attenuation to 10:1.
• Sets the trigger level (on the selected source channel) to -300 mV.
• Turns on trigger Noise Reject.
• Turns on Serial Decode.
• Sets the trigger to the specified serial bus (n of SBUS<n>).
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":SBUS<n>:FLEXray:TRIGger" on page 758
• ":SBUS<n>:FLEXray:BAUDrate" on page 751
• ":TRIGger[:EDGE]:LEVel" on page 1033
• ":SBUS<n>:FLEXray:SOURce" on page 757