Keysight InfiniiVision 3000T X-Series Oscilloscopes
Programmer's Guide
11 :COUNter Commands
When the optional DSOXDVMCTR digital voltmeter and counter analysis feature is
licensed, these commands control the counter feature. See "Introduction to
:COUNter Commands" on page 288.
Table 89 :COUNter Commands Summary
Command Query Options and Query Returns
n/a :COUNter:CURRent?
(see page 289)
<value> ::= current counter value
in NR3 format
:COUNter:ENABle {{0 |
OFF} | {1 | ON}} (see
page 290)
:COUNter:ENABle? (see
page 290)
{0 | 1}
:COUNter:MODE <mode>
(see page 291)
:COUNter:MODE (see
page 291)
<mode> ::= {FREQuency | PERiod |
TOTalize}
:COUNter:NDIGits
<value> (see page 292)
:COUNter:NDIGits (see
page 292)
<value> ::= 3 to 8 in NR1 format
:COUNter:SOURce
<source> (see
page 293)
:COUNter:SOURce? (see
page 293)
<source> ::= {CHANnel<n> |
TQEVent}
<n> ::= 1 to (# analog channels)
in NR1 format
:COUNter:TOTalize:CLE
ar (see page 294)
n/a n/a
:COUNter:TOTalize:GAT
E:ENABle {{0 | OFF} |
{1 | ON}} (see
page 295)
:COUNter:TOTalize:GAT
E:ENABle? (see
page 295)
{0 | 1}
:COUNter:TOTalize:GAT
E:POLarity <polarity>
(see page 296)
:COUNter:TOTalize:GAT
E:POLarity? (see
page 296)
<polarity> ::= {{NEGative |
FALLing} | {POSitive | RISing}}