462 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
22 :MEASure Commands
:MEASure:BRATe
(see page 1276)
Command Syntax
:MEASure:BRATe [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion<m> | MATH<m>
| WMEMory<r>}
<digital channels> ::= DIGital<d> for the MSO models
<d> ::= 0 to (# digital channels - 1) in NR1 format
<n> ::= 1 to (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:BRATe command installs a screen measurement and starts the bit
rate measurement. If the optional source parameter is specified, the currently
specified source is modified.
Query Syntax
:MEASure:BRATe? [<source>]
The :MEASure:BRATe? query measures all positive and negative pulse widths on
the waveform, takes the minimum value found of either width type and inverts that
minimum width to give a value in Hertz.
Return Format
<value><NL>
<value> ::= the bit rate value in Hertz
See Also "Introduction to :MEASure Commands" on page 458
":MEASure:SOURce" on page 499
":MEASure:FREQuency" on page 480
":MEASure:PERiod" on page 488
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).