:COUNter Commands 11
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 297
:COUNter:TOTalize:GATE:SOURce
(see page 1276)
Command Syntax
:COUNter:TOTalize:GATE:SOURce <source>
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :COUNter:TOTalize:GATE:SOURce command selects the analog channel that
has the totalizer gating signal.
Query Syntax
:COUNter:TOTalize:GATE:SOURce?
The :COUNter:TOTalize:GATE:SOURce? query returns the current totalizer gating
signal source.
Return Format
<source><NL>
See Also • ":COUNter:CURRent" on page 289
• ":COUNter:ENABle" on page 290
• ":COUNter:MODE" on page 291
• ":COUNter:NDIGits" on page 292
• ":COUNter:SOURce" on page 293
• ":COUNter:TOTalize:CLEar" on page 294
• ":COUNter:TOTalize:GATE:ENABle" on page 295
• ":COUNter:TOTalize:GATE:POLarity" on page 296
• ":COUNter:TOTalize:SLOPe" on page 298