:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 881
:SBUS<n>:UART:TRIGger:BASE
(see page 1276)
Command Syntax
:SBUS<n>:UART:TRIGger:BASE <base>
<base> ::= {ASCii | HEX}
The :SBUS<n>:UART:TRIGger:BASE command sets the front panel UART/RS232
trigger setup data selection option:
ASCii — front panel data selection is from ASCII values.
HEX — front panel data selection is from hexadecimal values.
The :SBUS<n>:UART:TRIGger:BASE setting does not affect the
:SBUS<n>:UART:TRIGger:DATA command which can always set data values using
ASCII or hexadecimal values.
Query Syntax
:SBUS<n>:UART:TRIGger:BASE?
The :SBUS<n>:UART:TRIGger:BASE? query returns the current UART base setting.
Return Format
<base><NL>
<base> ::= {ASC | HEX}
See Also "Introduction to :TRIGger Commands" on page 1007
":TRIGger:MODE" on page 1016
":SBUS<n>:UART:TRIGger:DATA" on page 883
NOTE
The :SBUS<n>:UART:TRIGger:BASE command is independent of the :SBUS<n>:UART:BASE
command which affects decode and Lister only.