:MEASure Power Commands 23
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 539
:MEASure:OPOWer
(see page 1276)
Command Syntax
:MEASure:OPOWer
The :MEASure:OPOWer command installs an output power measurement on
screen.
Before sending this command or query, you must specify the channels probing the
input voltage, input current, output voltage, and output current (using the
:POWer:SIGNals:SOURce:VOLTage<i> and :POWer:SIGNals:SOURce:CURRent<i>
commands) and you must perform the automated signals setup (using the
:POWer:SIGNals:AUTosetup EFFiciency command).
Query Syntax
:MEASure:OPOWer?
The :MEASure:OPOWer query returns the measured output power.
Return Format
<return_value><NL>
<return_value> ::= the output power value in NR3 format
See Also • ":POWer:SIGNals:SOURce:VOLTage<i>" on page 642
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 641
• ":POWer:SIGNals:AUTosetup" on page 624
• ":POWer:EFFiciency:APPLy" on page 595