EasyManuals Logo

Keysight Technologies InfiniiVision 3000T X Series User Manual

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #793 background imageLoading...
Page #793 background image
:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 793
:SBUS<n>:IIC:TRIGger:QUALifier
(see page 1276)
Command Syntax
:SBUS<n>:IIC:TRIGger:QUALifier <value>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
The :SBUS<n>:IIC:TRIGger:QUALifier command sets the IIC data qualifier when
TRIGger:IIC:TRIGger[:TYPE] is set to READEprom.
Query Syntax
:SBUS<n>:IIC:TRIGger:QUALifier?
The :SBUS<n>:IIC:TRIGger:QUALifier? query returns the current IIC data qualifier
value.
Return Format
<value><NL>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:MODE" on page 1016
• ":SBUS<n>:IIC:TRIGger[:TYPE]" on page 794