914 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
30 :SEARch Commands
:SEARch:RUNT:TIME
(see page 1276)
Command Syntax
:SEARch:RUNT:TIME <time>[suffix]
<time> ::= floating-point number in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
When searching for runt pulses whose widths are greater than or less than a time
(see :SEARch:RUNT:QUALifier), the :SEARch:RUNT:TIME command specifies the
time value.
Query Syntax
:SEARch:RUNT:TIME?
The :SEARch:RUNT:TIME? query returns the currently specified runt time value.
Return Format
<time><NL>
<time> ::= floating-point number in NR3 format
See Also • Chapter 30, “:SEARch Commands,” starting on page 889
• ":SEARch:RUNT:QUALifier" on page 912