954 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
30 :SEARch Commands
:SEARch:SERial:IIC:QUALifier
(see page 1276)
Command Syntax
:SEARch:SERial:IIC:QUALifier <value>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
The :SEARch:SERial:IIC:QUALifier command specifies the data value qualifier used
when searching for IIC EEPROM data read events.
Query Syntax
:SEARch:SERial:IIC:QUALifier?
The :SEARch:SERial:IIC:QUALifier? query returns the current data value qualifier
setting.
Return Format
<value><NL>
<value> ::= {EQU | NOT | LESS | GRE}
See Also Chapter 30, “:SEARch Commands,” starting on page 889
":SEARch:SERial:IIC:MODE" on page 949
":SEARch:SERial:IIC:PATTern:DATA" on page 952