:POWer Commands 26
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 619
:POWer:PSRR:FREQuency:MAXimum
(see page 1276)
Command Syntax
:POWer:PSRR:FREQuency:MAXimum <value>[suffix]
<value> ::= {10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000
| 20000000}
[suffix] ::= {Hz | kHz| MHz}
The :POWer:PSRR:FREQuency:MAXimum command sets the end sweep frequency
value. The PSRR measurement is displayed on a log scale, so you can select from
decade values in addition to the maximum frequency of 20 MHz.
Query Syntax
:POWer:PSRR:FREQuency:MAXimum?
The :POWer:PSRR:FREQuency:MAXimum query returns the maximum sweep
frequency setting.
Return Format
<value><NL>
<value> ::= {10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000
| 20000000}
See Also • ":POWer:PSRR:APPLy" on page 618
• ":POWer:PSRR:FREQuency:MINimum" on page 620
• ":POWer:PSRR:RMAXimum" on page 621