:TRIGger Commands 33
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1039
:TRIGger:GLITch:GREaterthan
(see page 1276)
Command Syntax
:TRIGger:GLITch:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= floating-point number in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:GREaterthan command sets the minimum pulse width
duration for the selected :TRIGger:GLITch:SOURce.
Query Syntax
:TRIGger:GLITch:GREaterthan?
The :TRIGger:GLITch:GREaterthan? query returns the minimum pulse width
duration time for :TRIGger:GLITch:SOURce.
Return Format
<greater_than_time><NL>
<greater_than_time> ::= floating-point number in NR3 format.
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:GLITch:SOURce" on page 1045
• ":TRIGger:GLITch:QUALifier" on page 1043
• ":TRIGger:MODE" on page 1016