830 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:SENT:SIGNal<s>:LENGth
(see page 1276)
Command Syntax
:SBUS<n>:SENT:SIGNal<s>:LENGth <length>
<s> ::= 1-6, in NR1 format.
<length> ::= from 1-24, in NR1 format.
The :SBUS<n>:SENT:SIGNal<s>:LENGth command specifies the bit length of the
signal being defined.
Fast Signal definition examples:
Query Syntax
:SBUS<n>:SENT:SIGNal<s>:LENGth?
The :SBUS<n>:SENT:SIGNal<s>:LENGth? query returns the signal bit length
setting.
Return Format
<length><NL>
<length> ::= from 1-24, in NR1 format.
See Also • ":SBUS<n>:SENT:CLOCk" on page 820
• ":SBUS<n>:SENT:CRC" on page 821
• ":SBUS<n>:SENT:DISPlay" on page 822
• ":SBUS<n>:SENT:FORMat" on page 824
• ":SBUS<n>:SENT:IDLE" on page 826
• ":SBUS<n>:SENT:LENGth" on page 827
• ":SBUS<n>:SENT:PPULse" on page 828
• ":SBUS<n>:SENT:SIGNal<s>:DISPlay" on page 829
• ":SBUS<n>:SENT:SIGNal<s>:MULTiplier" on page 832

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals