342 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
16 :EXTernal Trigger Commands
:EXTernal:RANGe
(see page 1276)
Command Syntax
:EXTernal:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
The :EXTernal:RANGe command is provided for product compatibility. When using
1:1 probe attenuation, the range can only be set to 8.0 V.
If the probe attenuation is changed, the range value is multiplied by the probe
attenuation factor.
Query Syntax
:EXTernal:RANGe?
The :EXTernal:RANGe? query returns the current full-scale range setting for the
external trigger.
Return Format
<range_argument><NL>
<range_argument> ::= external trigger range value in NR3 format
See Also "Introduction to :EXTernal Trigger Commands" on page 339
":EXTernal:PROBe" on page 341
"Introduction to :TRIGger Commands" on page 1007