Obsolete and Discontinued Commands 37
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1205
:MEASure:THResholds
(see page 1276)
Command Syntax
:MEASure:THResholds {T1090 | T2080 | VOLTage}
The :MEASure:THResholds command selects the thresholds used when making
time measurements.
Query Syntax
:MEASure:THResholds?
The :MEASure:THResholds? query returns the current thresholds selected when
making time measurements.
Return Format
{T1090 | T2080 | VOLTage}<NL>
{T1090} uses the 10% and 90% levels of the selected waveform.
{T2080} uses the 20% and 80% levels of the selected waveform.
{VOLTage} uses the upper and lower voltage thresholds set by the
UPPer and LOWer commands on the selected waveform.
See Also "Introduction to :MEASure Commands" on page 458
":MEASure:LOWer" on page 1202
":MEASure:UPPer" on page 1211
NOTE
The :MEASure:THResholds command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 467)
instead.