:TRIGger Commands 33
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1033
:TRIGger[:EDGE]:LEVel
(see page 1276)
Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= ±(external range setting) in NR3 format
for external triggers
<level> ::= ±8 V for digital channels (MSO models)
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d> | EXTernal}
for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the active
trigger source.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current trigger
source.
Return Format
<level><NL>
See Also "Introduction to :TRIGger Commands" on page 1007
":TRIGger[:EDGE]:SOURce" on page 1036
":EXTernal:RANGe" on page 342
":POD<n>:THReshold" on page 587
":DIGital<d>:THReshold" on page 313
NOTE
If the optional source is specified and is not the active source, the level on the active source is
not affected and the active source is not changed.