37 Obsolete and Discontinued Commands
:FUNCtion:VIEW
(see page 1276)
Command Syntax
:FUNCtion:VIEW <view>
<view> ::= {{1 | ON} | (0 | OFF}}
The :FUNCtion:VIEW command turns the selected function on or off. When ON is
selected, the function performs as specified using the other FUNCtion commands.
When OFF is selected, function is neither calculated nor displayed.
Query Syntax
:FUNCtion:VIEW?
The :FUNCtion:VIEW? query returns the current state of the selected function.
Return Format
<view><NL>
<view> ::= {1 | 0}
NOTE
The :FUNCtion:VIEW command is provided for backward compatibility to previous
oscilloscopes. Use the :FUNCtion:DISPlay command (see page 377) instead.