646 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:SWITch:CONDuction
(see page 1276)
Command Syntax
:POWer:SWITch:CONDuction <conduction>
<conduction> ::= {WAVeform | RDS | VCE}
The :POWer:SWITch:CONDuction command specifies the conduction calculation
method:
WAVeform — The Power waveform uses the original voltage waveform data,
and the calculation is: P = V x I
RDS — Rds(on) — The Power waveform includes error correction:
In the On Zone (where the voltage level is below V Ref) – the Power
calculation is: P = Id2 x Rds(on)
Specify Rds(on) using the :POWer:SWITch:RDS command.
In the Off Zone (where the current level is below I Ref) – the Power
calculation is: P = 0 Watt.
VCE — Vce(sat) — The Power waveform includes error correction:
In the On Zone (where the voltage level is below V Ref) – the Power
calculation is: P = Vce(sat) x Ic
Specify Vce(sat) using the :POWer:SWITch:VCE command.
In the Off Zone (where the current level is below I Ref) – the Power
calculation is: P = 0 Watt.
Query Syntax
:POWer:SWITch:CONDuction?
The :POWer:SWITch:CONDuction query returns the conduction calculation
method.
Return Format
<conduction><NL>
<conduction> ::= {WAV | RDS | VCE}
See Also ":POWer:SWITch:APPLy" on page 645
":POWer:SWITch:IREFerence" on page 647
":POWer:SWITch:RDS" on page 648
":POWer:SWITch:VCE" on page 649
":POWer:SWITch:VREFerence" on page 650