EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :FUNCtion<m>:RANGe

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
394 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
18 :FUNCtion<m> Commands
:FUNCtion<m>:RANGe
(see page 1276)
Command Syntax
:FUNCtion<m>:RANGe <range>
<m> ::= 1 to (# math functions) in NR1 format
<range> ::= the full-scale vertical axis value in NR3 format.
The :FUNCtion<m>:RANGe command defines the full-scale vertical axis for the
selected function.
Query Syntax
:FUNCtion<m>:RANGe?
The :FUNCtion<m>:RANGe? query returns the current full-scale range value for
the selected function.
Return Format
<range><NL>
<range> ::= the full-scale vertical axis value in NR3 format.
See Also "Introduction to :FUNCtion<m> Commands" on page 368