876 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:UART:FRAMing
(see page 1276)
Command Syntax
:SBUS<n>:UART:FRAMing <value>
<value> ::= {OFF | <decimal> | <nondecimal>}
<decimal> ::= 8-bit integer in decimal from 0-255 (0x00-0xff)
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
The :SBUS<n>:UART:FRAMing command determines the byte value to use for
framing (end of packet) or to turn off framing for UART decode.
Query Syntax
:SBUS<n>:UART:FRAMing?
The :SBUS<n>:UART:FRAMing? query returns the current UART decode base
setting.
Return Format
<value><NL>
<value> ::= {OFF | <decimal>}
<decimal> ::= 8-bit integer in decimal from 0-255
Errors • "-241, Hardware missing" on page 1235
See Also • "Introduction to :SBUS<n> Commands" on page 697
• ":SBUS<n>:UART Commands" on page 866

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals