EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :SEARch:PEAK:EXCursion

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
906 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
30 :SEARch Commands
:SEARch:PEAK:EXCursion
(see page 1276)
Command Syntax
:SEARch:PEAK:EXCursion <delta_level>
<delta_level> ::= required change in level to be recognized as a peak,
in NR3 format.
The :SEARch:PEAK:EXCursion command specifies the change in level that must
occur (in other words, hysteresis) to be recognized as a peak.
The threshold level units are specified by the :FUNCtion<m>[:FFT]:VTYPe
command.
Query Syntax
:SEARch:PEAK:EXCursion?
The :SEARch:PEAK:EXCursion? query returns the specified excursion delta level
value.
Return Format
<delta_level><NL>
<delta_level> ::= in NR3 format.
See Also ":FUNCtion<m>[:FFT]:VTYPe" on page 382
":SEARch:PEAK:NPEaks" on page 907
":SEARch:PEAK:SOURce" on page 908