1204 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
37 Obsolete and Discontinued Commands
:MEASure:TDELta
(see page 1276)
Query Syntax
:MEASure:TDELta?
The :MEASure:TDELta? query returns the time difference between the Tstop
marker (X2 cursor) and the Tstart marker (X1 cursor).
Tdelta = Tstop - Tstart
Tstart is the time at the start marker (X1 cursor) and Tstop is the time at the stop
marker (X2 cursor). No measurement is made when the :MEASure:TDELta? query
is received by the oscilloscope. The delta time value that is output is the current
value. This is the same value as the front-panel cursors delta X value.
Return Format
<value><NL>
<value> ::= time difference between start and stop markers in NR3 format
See Also • "Introduction to :MARKer Commands" on page 426
• "Introduction to :MEASure Commands" on page 458
• ":MARKer:X1Position" on page 430
• ":MARKer:X2Position" on page 432
• ":MARKer:XDELta" on page 434
• ":MEASure:TSTArt" on page 1208
• ":MEASure:TSTOp" on page 1209
NOTE
The :MEASure:TDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:XDELta command (see page 434) instead.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals