:EXTernal Trigger Commands 16
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 341
:EXTernal:PROBe
(see page 1276)
Command Syntax
:EXTernal:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
The :EXTernal:PROBe command specifies the probe attenuation factor for the
external trigger. The probe attenuation factor may be 0.1 to 1000. This command
does not change the actual input sensitivity of the oscilloscope. It changes the
reference constants for scaling the display factors and for setting trigger levels.
If an AutoProbe probe is connected to the oscilloscope, the attenuation value
cannot be changed from the sensed value. Attempting to set the oscilloscope to
an attenuation value other than the sensed value produces an error.
Query Syntax
:EXTernal:PROBe?
The :EXTernal:PROBe? query returns the current probe attenuation factor for the
external trigger.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also • "Introduction to :EXTernal Trigger Commands" on page 339
• ":EXTernal:RANGe" on page 342
• "Introduction to :TRIGger Commands" on page 1007
• ":CHANnel<n>:PROBe" on page 275

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals