1144 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
35 :WGEN<w> Commands
:WGEN<w>:MODulation:AM:DEPTh
(see page 1276)
Command Syntax
:WGEN<w>:MODulation:AM:DEPTh <percent>
<w> ::= 1 in NR1 format
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN<w>:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used by the
modulation. For example, a depth setting of 80% causes the output amplitude to
vary from 10% to 90% (90% – 10% = 80%) of the original amplitude as the
modulating signal goes from its minimum to maximum amplitude.
Query Syntax
:WGEN<w>:MODulation:AM:DEPTh?
The :WGEN<w>:MODulation:AM:DEPTh? query returns the AM depth percentage
setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also • ":WGEN<w>:MODulation:AM:FREQuency" on page 1145
• ":WGEN<w>:MODulation:FM:DEViation" on page 1146
• ":WGEN<w>:MODulation:FM:FREQuency" on page 1147
• ":WGEN<w>:MODulation:FSKey:FREQuency" on page 1148
• ":WGEN<w>:MODulation:FSKey:RATE" on page 1149
• ":WGEN<w>:MODulation:FUNCtion" on page 1150
• ":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry" on page 1151
• ":WGEN<w>:MODulation:STATe" on page 1153
• ":WGEN<w>:MODulation:TYPE" on page 1154