EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :WGEN<w>:MODulation:AM:DEPTh

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1144 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
35 :WGEN<w> Commands
:WGEN<w>:MODulation:AM:DEPTh
(see page 1276)
Command Syntax
:WGEN<w>:MODulation:AM:DEPTh <percent>
<w> ::= 1 in NR1 format
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN<w>:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used by the
modulation. For example, a depth setting of 80% causes the output amplitude to
vary from 10% to 90% (90% – 10% = 80%) of the original amplitude as the
modulating signal goes from its minimum to maximum amplitude.
Query Syntax
:WGEN<w>:MODulation:AM:DEPTh?
The :WGEN<w>:MODulation:AM:DEPTh? query returns the AM depth percentage
setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also ":WGEN<w>:MODulation:AM:FREQuency" on page 1145
":WGEN<w>:MODulation:FM:DEViation" on page 1146
":WGEN<w>:MODulation:FM:FREQuency" on page 1147
":WGEN<w>:MODulation:FSKey:FREQuency" on page 1148
":WGEN<w>:MODulation:FSKey:RATE" on page 1149
":WGEN<w>:MODulation:FUNCtion" on page 1150
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry" on page 1151
":WGEN<w>:MODulation:STATe" on page 1153
":WGEN<w>:MODulation:TYPE" on page 1154

Table of Contents