184 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
5 Common (*) Commands
*SRE (Service Request Enable)
(see page 1276)
Command Syntax
*SRE <mask>
<mask> ::= integer with values defined in the following table.
The *SRE command sets the bits in the Service Request Enable Register. The
Service Request Enable Register contains a mask value for the bits to be enabled
in the Status Byte Register. A one in the Service Request Enable Register enables
the corresponding bit in the Status Byte Register. A zero disables the bit.
From
Op er a t i o n
Status
Registers
10234567
+
OPER TRGMAVESB
RQS/
MSS
TRG
Reg
TER? Trigger Event Register
*SRE
*SRE?
*STB? Status Byte Register
Service Request Enable
(Mask) Register
Ou t p u t
Queue
SRQ
From
Standard
Event
Status
Registers
USR
MSG
Service Request
OR

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals