:POWer Commands 26
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 653
:POWer:TRANsient:IINitial
(see page 1276)
Command Syntax
:POWer:TRANsient:IINitial <value>[suffix]
<value> ::= Initial current value in NR3 format
[suffix] ::= {A | mA}
The :POWer:TRANsient:IINitial command to specify the initial load current value.
The initial load current will be used as a reference and to trigger the oscilloscope.
Query Syntax
:POWer:TRANsient:IINitial?
The :POWer:TRANsient:IINitial query returns the initial load current value.
Return Format
<value><NL>
<value> ::= Initial current value in NR3 format
See Also • ":POWer:SIGNals:VSTeady:TRANsient" on page 640
• ":POWer:TRANsient:APPLy" on page 651
• ":POWer:TRANsient:EXIT" on page 652
• ":POWer:TRANsient:INEW" on page 654
• ":POWer:TRANsient:NEXT" on page 655