:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 741
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth
(see page 1276)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command sets the number of
8-bit bytes in the CAN data string. The number of bytes in the string can be
anywhere from 1 bytes to 8 bytes (64 bits). The value for these bytes is set by the
:SBUS<n>:CAN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth? query returns the current
CAN data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors • "-241, Hardware missing" on page 1235
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 739
• ":SBUS<n>:CAN:SOURce" on page 734