854 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:SPI:CLOCk:TIMeout
(see page 1276)
Command Syntax
:SBUS<n>:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR3 format
The :SBUS<n>:SPI:CLOCk:TIMeout command sets the SPI signal clock timeout
resource in seconds from 100 ns to 10 s when the :SBUS<n>:SPI:FRAMing
command is set to TIMeout. The timer is used to frame a signal by a clock timeout.
Query Syntax
:SBUS<n>:SPI:CLOCk:TIMeout?
The :SBUS<n>:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":SBUS<n>:SPI:CLOCk:SLOPe" on page 853
• ":SBUS<n>:SPI:SOURce:CLOCk" on page 856
• ":SBUS<n>:SPI:FRAMing" on page 855