:TRIGger Commands 33
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1015
:TRIGger:LEVel:LOW
(see page 1276)
Command Syntax
:TRIGger:LEVel:LOW <level>, <source>
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:LEVel:LOW command sets the low trigger voltage level voltage for
the specified source.
High and low trigger levels are used with runt triggers and rise/fall time
(transition) triggers.
Query Syntax
:TRIGger:LEVel:LOW? <source>
The :TRIGger:LEVel:LOW? query returns the low trigger voltage level for the
specified source.
Return Format
<level><NL>
See Also "Introduction to :TRIGger Commands" on page 1007
":TRIGger:LEVel:HIGH" on page 1014
":TRIGger:RUNT Commands" on page 1056
":TRIGger:TRANsition Commands" on page 1067
":TRIGger[:EDGE]:SOURce" on page 1036

Table of Contents