:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 801
:SBUS<n>:LIN:SOURce
(see page 1276)
Command Syntax
:SBUS<n>:LIN:SOURce <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d>} for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :SBUS<n>:LIN:SOURce command sets the source for the LIN signal.
Query Syntax
:SBUS<n>:LIN:SOURce?
The :SBUS<n>:LIN:SOURce? query returns the current source for the LIN signal.
Return Format
<source><NL>
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:MODE" on page 1016
• ":SBUS<n>:LIN:TRIGger" on page 804
• ":SBUS<n>:LIN:SIGNal:DEFinition" on page 1226