EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :FFT:CLEar

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:FFT Commands 17
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 349
:FFT:CLEar
(see page 1276)
Command Syntax
:FFT:CLEar
When the FFT display mode is AVERage, MAXHold, or MINHold, the :FFT:CLEar
command clears the number of evaluated waveforms.
See Also ":FFT:AVERage:COUNt" on page 347
":FFT:CENTer" on page 348
":FFT:DISPlay" on page 350
":FFT:DMODe" on page 351
":FFT:OFFSet" on page 353
":FFT:RANGe" on page 354
":FFT:REFerence" on page 355
":FFT:SCALe" on page 356
":FFT:SOURce1" on page 357
":FFT:SPAN" on page 358
":FFT:STARt" on page 359
":FFT:STOP" on page 360
":FFT:VTYPe" on page 361
":FFT:WINDow" on page 362

Table of Contents

Related product manuals