862 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA
(see page 1276)
Command Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA <string>
<string> ::= "nn...n" where n ::= {0 |1|X|$}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA command defines the SPI data
pattern resource according to the string parameter. This pattern, along with the
data width, control the data pattern searched for in the data stream.
If the string parameter starts with "0x", it is a hexadecimal string made up of
hexadecimal and X (don't care) characters; otherwise, it is a binary string made up
of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA?
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA? query returns the current
settings of the specified SPI data pattern resource in the binary string format.
Return Format
<string><NL>
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh" on page 863
• ":SBUS<n>:SPI:SOURce:MOSI" on page 859
NOTE
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh should be set before
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals