:SEARch Commands 30
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 947
:SEARch:SERial:I2S:RANGe
(see page 1276)
Command Syntax
:SEARch:SERial:I2S:RANGe <lower>, <upper>
<lower> ::= 32-bit integer in signed decimal, <nondecimal>, or <string>
<upper> ::= 32-bit integer in signed decimal, <nondecimal>, or <string>
<nondecimal> ::= #Hnn...n where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :SEARch:SERial:I2S:RANGe command specifies the data value range when
searching for I2S events in the INRange and OUTRange search modes (set by the
:SEARch:SERial:I2S:MODE command).
You can enter the parameters in any order — the smaller value becomes the
<lower> and the larger value becomes the <upper>.
Query Syntax
:SEARch:SERial:I2S:RANGe?
The :SEARch:SERial:I2S:RANGe? query returns the current data value range
setting.
Return Format
<lower>, <upper><NL>
<lower> ::= 32-bit integer in signed decimal
<upper> ::= 32-bit integer in signed decimal
See Also Chapter 30, “:SEARch Commands,” starting on page 889
":SEARch:SERial:I2S:MODE" on page 944