1124 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
34 :WAVeform Commands
:WAVeform:YREFerence
(see page 1276)
Query Syntax
:WAVeform:YREFerence?
The :WAVeform:YREFerence? query returns the y-reference value for the currently
specified source. This value specifies the data point value where the y-origin
occurs. In this product, this is the data point value of the center of the screen. It is
undefined if the format is ASCii.
Return Format
<value><NL>
<value> ::= y-reference value in the current preamble in 32-bit
NR1 format
See Also • "Introduction to :WAVeform Commands" on page 1091
• ":WAVeform:PREamble" on page 1106
• ":WAVeform:YORigin" on page 1123
Example Code • "Example Code" on page 1107