Keysight Technologies InfiniiVision 3000T X Series - Mask Test Event Event Register (:MTERegister[:EVENt])

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Status Reporting 39
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1261
Mask Test Event Event Register (:MTERegister[:EVENt])
The Mask Test Event Event Register register hosts these bits:
The :MTERegister[:EVENt]? query returns the value of, and clears, the Mask Test
Event Event Register.
Name Location Description
Complete bit 0 Is set when the mask test is complete.
Fail bit 1 Is set when there is a mask test failure.
Started bit 8 Is set when mask testing is started.
Auto Mask bit 10 Is set when auto mask creation is completed.