:CHANnel<n> Commands 10
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 281
:CHANnel<n>:PROTection
(see page 1276)
Command Syntax
:CHANnel<n>:PROTection[:CLEar]
<n> ::= 1 to (# analog channels) in NR1 format| 4}
When the analog channel input impedance is set to 50Ω, the input channels are
protected against overvoltage. When an overvoltage condition is sensed, the input
impedance for the channel is automatically changed to 1 M
Ω.
The :CHANnel<n>:PROTection[:CLEar] command is used to clear (reset) the
overload protection. It allows the channel to be used again in 50
Ω mode after the
signal that caused the overload has been removed from the channel input.
Reset the analog channel input impedance to 50
Ω (see
":CHANnel<n>:IMPedance" on page 271) after clearing the overvoltage
protection.
Query Syntax
:CHANnel<n>:PROTection?
The :CHANnel<n>:PROTection query returns the state of the input protection for
CHANnel<n>. If a channel input has experienced an overload, TRIP (tripped) will
be returned; otherwise NORM (normal) is returned.
Return Format
{NORM | TRIP}<NL>
See Also • "Introduction to :CHANnel<n> Commands" on page 267
• ":CHANnel<n>:COUPling" on page 269
• ":CHANnel<n>:IMPedance" on page 271
• ":CHANnel<n>:PROBe" on page 275

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals