258 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
9 :CALibrate Commands
:CALibrate:LABel
(see page 1276)
Command Syntax
:CALibrate:LABel <string>
<string> ::= quoted ASCII string of up to 32 characters in length,
not including the quotes
The CALibrate:LABel command saves a string that is up to 32 characters in length
into the instrument's non-volatile memory. The string may be used to record
calibration dates or other information as needed.
Query Syntax
:CALibrate:LABel?
The :CALibrate:LABel? query returns the contents of the calibration label string.
Return Format
<string><NL>
<string>::= unquoted ASCII string of up to 32 characters in length
See Also "Introduction to :CALibrate Commands" on page 256