348 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
17 :FFT Commands
:FFT:CENTer
(see page 1276)
Command Syntax
:FFT:CENTer <frequency>
<frequency> ::= the current center frequency in NR3 format. The range
of legal values is from 0 Hz to 25 GHz.
The :FFT:CENTer command sets the center frequency when FFT (Fast Fourier
Transform) is selected.
Query Syntax
:FFT:CENTer?
The :FFT:CENTer? query returns the current center frequency in Hertz.
Return Format
<frequency><NL>
<frequency> ::= the current center frequency in NR3 format. The range
of legal values is from 0 Hz to 25 GHz.
See Also ":FFT:AVERage:COUNt" on page 347
":FFT:CENTer" on page 348
":FFT:DISPlay" on page 350
":FFT:DMODe" on page 351
":FFT:OFFSet" on page 353
":FFT:RANGe" on page 354
":FFT:REFerence" on page 355
":FFT:SCALe" on page 356
":FFT:SOURce1" on page 357
":FFT:SPAN" on page 358
":FFT:STARt" on page 359
":FFT:STOP" on page 360
":FFT:VTYPe" on page 361
":FFT:WINDow" on page 362
NOTE
After a *RST (Reset) or :AUToscale command, the values returned by the :FFT:CENTer? and
:FFT:SPAN? queries depend on the current :TIMebase:RANGe value. Once you change either
the :FFT:CENTer or :FFT:SPAN value, they no longer track the :TIMebase:RANGe value.