EasyManuals Logo

Keysight Technologies InfiniiVision 3000T X Series User Manual

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1058 background imageLoading...
Page #1058 background image
1058 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
:TRIGger:RUNT:QUALifier
(see page 1276)
Command Syntax
:TRIGger:RUNT:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | NONE}
The :TRIGger:RUNT:QUALifier command selects the qualifier used for specifying
runt pulse widths:
• GREaterthan — triggers on runt pulses whose width is greater than the
:TRIGger:RUNT:TIME.
• LESSthan — triggers on runt pulses whose width is less than the
:TRIGger:RUNT:TIME.
• NONE — triggers on runt pulses of any width.
Query Syntax
:TRIGger:RUNT:QUALifier?
The :TRIGger:RUNT:QUALifier? query returns the runt trigger qualifier setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS NONE}
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:MODE" on page 1016
• ":TRIGger:RUNT:TIME" on page 1060