:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 883
:SBUS<n>:UART:TRIGger:DATA
(see page 1276)
Command Syntax
:SBUS<n>:UART:TRIGger:DATA <value>
<value> ::= 8-bit integer from 0-255 (0x00-0xff) in decimal,
<hexadecimal>, <binary>, or <quoted_string> format
<hexadecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<binary> ::= #Bnn...n where n ::= {0 | 1} for binary
<quoted_string> ::= any of the 128 valid 7-bit ASCII characters
(or standard abbreviations)
The :SBUS<n>:UART:TRIGger:DATA command selects the data byte value (0x00 to
0xFF) for the trigger QUALifier when in UART mode. The data value is used when
one of the RD or TD trigger types is selected.
When entering an ASCII character via the quoted string, it must be one of the 128
valid characters (case-sensitive): "NUL", "SOH", "STX", "ETX", "EOT", "ENQ",
"ACK", "BEL", "BS", "HT", "LF", "VT", "FF", "CR", "SO","SI", "DLE", "DC1", "DC2",
"DC3", "DC4", "NAK", "SYN", "ETB", "CAN", "EM", "SUB", "ESC", "FS","GS", "RS",
"US", "SP", "!", "\"", "#", "$", "%","&", "\'", "(", ")", "*", "+", ",", "-", ".", "/","0", "1", "2",
"3", "4", "5", "6", "7", "8", "9",":", ";", "<", "=", ">", "?", "@", "A", "B", "C","D", "E", "F",
"G", "H", "I", "J", "K", "L", "M","N", "O", "P", "Q", "R", "S", "T", "U", "V", "W", "X", "Y",
"Z","[", "\\", "]", "^", "_", "`", "a", "b", "c", "d", "e", "f", "g", "h", "i", "j", "k", "l", "m", "n",
"o", "p", "q", "r", "s", "t", "u", "v", "w", "x", "y", "z", "{", "|", "}", "~", or "DEL".
Query Syntax
:SBUS<n>:UART:TRIGger:DATA?
The :SBUS<n>:UART:TRIGger:DATA? query returns the current UART trigger data
value.
Return Format
<value><NL>
<value> ::= 8-bit integer in decimal from 0-255
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:MODE" on page 1016
• ":SBUS<n>:UART:TRIGger:BASE" on page 881
• ":SBUS<n>:UART:TRIGger:TYPE" on page 886