:WAVeform Commands 34
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1097
:WAVeform:BYTeorder
(see page 1276)
Command Syntax
:WAVeform:BYTeorder <value>
<value> ::= {LSBFirst | MSBFirst}
The :WAVeform:BYTeorder command sets the output sequence of the WORD data.
The parameter MSBFirst sets the most significant byte to be transmitted first. The
parameter LSBFirst sets the least significant byte to be transmitted first. This
command affects the transmitting sequence only when :WAVeform:FORMat
WORD is selected. The default setting is LSBFirst.
Query Syntax
:WAVeform:BYTeorder?
The :WAVeform:BYTeorder query returns the current output sequence.
Return Format
<value><NL>
<value> ::= {LSBF | MSBF}
See Also • "Introduction to :WAVeform Commands" on page 1091
• ":WAVeform:DATA" on page 1099
• ":WAVeform:FORMat" on page 1101
• ":WAVeform:PREamble" on page 1106
Example Code • "Example Code" on page 1112
• "Example Code" on page 1107

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals