:CALibrate Commands 9
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 261
:CALibrate:STARt
(see page 1276)
Command Syntax
:CALibrate:STARt
The CALibrate:STARt command starts the user calibration procedure.
See Also • "Introduction to :CALibrate Commands" on page 256
• ":CALibrate:PROTected" on page 260
NOTE
Before starting the user calibration procedure, you must set the rear panel CALIBRATION
switch to UNPROTECTED, and you must connect BNC cables from the TRIG OUT connector to
the analog channel inputs. See the User's Guide for details.