:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 705
:SBUS<n>:A429:BASE
(see page 1276)
Command Syntax
:SBUS<n>:A429:BASE <base>
<base> ::= {BINary | HEX}
The :SBUS<n>:A429:BASE command selects between hexadecimal and binary
display of the decoded data.
The BASE command has no effect on the SDI and SSM fields, which are always
displayed in binary, nor the Label field, which is always displayed in octal.
Query Syntax
:SBUS<n>:A429:BASE?
The :SBUS<n>:A429:BASE? query returns the current ARINC 429 base setting.
Return Format
<base><NL>
<base> ::= {BIN | HEX}
Errors • "-241, Hardware missing" on page 1235
See Also • "Introduction to :SBUS<n> Commands" on page 697
• ":SBUS<n>:MODE" on page 701
• ":SBUS<n>:A429:FORMat" on page 709