:MEASure Commands 22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 519
:MEASure:VTIMe
(see page 1276)
Query Syntax
:MEASure:VTIMe? <vtime_argument>[,<source>]
<vtime_argument> ::= time from trigger in seconds
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion<m> | MATH<m>
| WMEMory<r>}
<digital channels> ::= DIGital<d> for the MSO models
<d> ::= 0 to (# digital channels - 1) in NR1 format
<n> ::= 1 to (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:VTIMe? query returns the value at a specified time on the source
specified with :MEASure:SOURce. The specified time must be on the screen and is
referenced to the trigger event. If the optional source parameter is specified, the
current source is modified.
Return Format
<value><NL>
<value> ::= value at the specified time in NR3 format
See Also • "Introduction to :MEASure Commands" on page 458
• ":MEASure:SOURce" on page 499
• ":MEASure:TEDGe" on page 507
• ":MEASure:TVALue" on page 509
NOTE
This query is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals