:TRIGger Commands 33
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1079
:TRIGger:TV:UDTV:HSYNc
(see page 1276)
Command Syntax
:TRIGger:TV:UDTV:HSYNc {{0 | OFF} | {1 | ON}}
The :TRIGger:TV:UDTV:HSYNc command enables or disables the horizontal sync
control in the Generic video trigger.
For interleaved video, enabling the HSYNc control and setting the HTIMe
adjustment to the sync time of the probed video signal allows the ENUMber
function to count only lines and not double count during equalization.
Additionally, the Field Holdoff can be adjusted so that the oscilloscope triggers
once per frame.
Similarly, for progressive video with a tri-level sync, enabling the HSYNc control
and setting the HTIMe adjustment to the sync time of the probed video signal
allows the ENUMber function to count only lines and not double count during
vertical sync.
Query Syntax
:TRIGger:TV:UDTV:HSYNc?
The :TRIGger:TV:UDTV:HSYNc query returns the horizontal sync control setting.
Return Format
{0 | 1}
See Also • ":TRIGger:TV:STANdard" on page 1077
• ":TRIGger:TV:UDTV:HTIMe" on page 1080
• ":TRIGger:TV:UDTV:ENUMber" on page 1078
• ":TRIGger:TV:UDTV:PGTHan" on page 1081

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals