:MARKer Commands 21
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 429
:MARKer:MODE
(see page 1276)
Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual | WAVeform | BINary | HEX}
The :MARKer:MODE command sets the cursors mode:
• OFF — removes the cursor information from the display.
• MANual — enables manual placement of the X and Y cursors.
If the front-panel cursors are off, or are set to the front-panel Hex or Binary
mode, setting :MARKer:MODE MANual will put the cursors in the front-panel
Normal mode.
• MEASurement — cursors track the most recent measurement.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement source
(:MEASure:SOURce). Setting the measurement source remotely always sets the
marker sources.
• WAVeform — the Y1 cursor tracks the voltage value at the X1 cursor of the
waveform specified by the X1Y1source, and the Y2 cursor does the same for the
X2 cursor and its X2Y2source.
• BINary — logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed in the Cursor sidebar dialog in binary.
• HEX — logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed in the Cursor sidebar dialog in hexadecimal.
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN | WAV | BIN | HEX}
See Also • "Introduction to :MARKer Commands" on page 426
• ":MARKer:X1Y1source" on page 431
• ":MARKer:X2Y2source" on page 433
• ":MEASure:SOURce" on page 499
• ":MARKer:X1Position" on page 430
• ":MARKer:X2Position" on page 432
• ":MARKer:Y1Position" on page 437
• ":MARKer:Y2Position" on page 438

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals