312 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
13 :DIGital<d> Commands
:DIGital<d>:SIZE
(see page 1276)
Command Syntax
:DIGital<d>:SIZE <value>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<value> ::= {SMALl | MEDium | LARGe}
The :DIGital<d>:SIZE command specifies the size of digital channels on the
display. Sizes are set for all digital channels. Therefore, if you set the size on digital
channel 0 (for example), the same size is set on all other as well.
Query Syntax
:DIGital<d>:SIZE?
The :DIGital<d>:SIZE? query returns the size setting for the specified digital
channels.
Return Format
<size_value><NL>
<size_value> ::= {SMAL | MED | LARG}
See Also • "Introduction to :DIGital<d> Commands" on page 308
• ":POD<n>:SIZE" on page 586
• ":DIGital<d>:POSition" on page 311
NOTE
This command is only valid for the MSO models.