EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
766 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE
(see page 1276)
Command Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE <frame_type>
<frame_type> ::= {NORMal | STARtup | NULL | SYNC | NSTArtup | NNULl |
NSYNc | ALL}
The :SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE command sets the FlexRay frame
type to trigger on. The frame type setting is only valid when the FlexRay trigger
mode is set to FRAME.
NORMal — will trigger on only normal (NSTArtup & NNULl & NSYNc) frames.
STARtup — will trigger on only startup frames.
NULL — will trigger on only null frames.
SYNC — will trigger on only sync frames.
NSTArtup — will trigger on frames other than startup frames.
NNULl — will trigger on frames other than null frames.
NSYNc — will trigger on frames other than sync frames.
ALL — will trigger on all FlexRay frame types.
Query Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE?
The :SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE? query returns the current frame
type setting for the FlexRay frame trigger setup.
Return Format
<frame_type><NL>
<frame_type> ::= {NORM | STAR | NULL | SYNC | NSTA | NNUL |
NSYN | ALL}
See Also "Introduction to :TRIGger Commands" on page 1007
":TRIGger:MODE" on page 1016
":SBUS<n>:FLEXray:TRIGger" on page 758

Table of Contents