:WAVeform Commands 34
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1109
:WAVeform:SEGMented:COUNt
(see page 1276)
Query Syntax
:WAVeform:SEGMented:COUNt?
The :WAVeform:SEGMented:COUNt query returns the number of memory
segments in the acquired data. You can use the :WAVeform:SEGMented:COUNt?
query while segments are being acquired (although :DIGitize blocks subsequent
queries until the full segmented acquisition is complete).
The segmented memory acquisition mode is enabled with the :ACQuire:MODE
command. The number of segments to acquire is set using the
:ACQuire:SEGMented:COUNt command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands.
Return Format
<count> ::= an integer from 2 to 1000 in NR1 format (count set by
:ACQuire:SEGMented:COUNt).
See Also • ":ACQuire:MODE" on page 235
• ":ACQuire:SEGMented:COUNt" on page 238
• ":DIGitize" on page 203
• ":SINGle" on page 226
• ":RUN" on page 224
• "Introduction to :WAVeform Commands" on page 1091
Example Code • "Example Code" on page 239
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals