1248 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
39 Status Reporting
Service Request Enable Register (SRE)
Setting the Service Request Enable Register bits enable corresponding bits in the
Status Byte Register. These enabled bits can then set RQS and MSS (bit 6) in the
Status Byte Register.
Bits are set in the Service Request Enable Register using the *SRE command and
the bits that are set are read with the *SRE? query.
Example The following example sets bit 4 (MAV) and bit 5 (ESB) in the Service Request
Enable Register.
myScope.WriteString "*SRE " + CStr(CInt("&H30"))
This example uses the decimal parameter value of 48, the string returned by
CStr(CInt("&H30")), to enable the oscilloscope to generate an SRQ interrupt under
the following conditions:
• When one or more bytes in the Output Queue set bit 4 (MAV).
• When an enabled event in the Standard Event Status Register generates a
summary bit that sets bit 5 (ESB).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals