932 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
30 :SEARch Commands
:SEARch:SERial:CAN:PATTern:ID:MODE
(see page 1276)
Command Syntax
:SEARch:SERial:CAN:PATTern:ID:MODE <value>
<value> ::= {STANdard | EXTended}
The :SEARch:SERial:CAN:PATTern:ID:MODE command specifies whether a
standard ID value or an extended ID value is used when searching for a CAN event.
The ID value is specified using the :SEARch:SERial:CAN:PATTern:ID command.
Query Syntax
:SEARch:SERial:CAN:PATTern:ID:MODE?
The :SEARch:SERial:CAN:PATTern:ID:MODE? query returns the current setting.
Return Format
<value><NL>
<value> ::= {STAN | EXT}
See Also Chapter 30, “:SEARch Commands,” starting on page 889
":SEARch:SERial:CAN:MODE" on page 927
":SEARch:SERial:CAN:PATTern:ID" on page 931